12.Specifications - 12.1.General Specifications
1WMPD4004444
143
12.1.2. General Specifications
Item
Specifications
Power
Rated power voltage
Allowed range of variation
in power voltage
Rated power frequency
Allowed range of variation
in power frequency
Withstand voltage
Insulation resistance
Power consumption
Power fuse
AC 100 to 240 V
AC 90 to 264 V
50/60 Hz
47 to 63 Hz
Between power and case 1500 V AC for 1 minute
Between power and case
100 MΩ or more at 500
VDC
For printer recording
300 VA (maximum printing
state)
When recording is stopped 80 VA
For standby
5 VA (power cord connected
and power off)
Internal (not replaceable)
Locations for use
Indoor, Pollution Degree 2
*
1
, elevation 2000 m or lower
Warmup time
60 minutes or longer
Operating environment Temperature
humidity
0 to 40°C
35 to 85 RH% or less (without condensation)
Storage environment Temperature
humidity
20 to 60°C
20 to 85 RH% or less (without condensation)
Vibration resistance
Sine wave vibration
Vibration frequency
Vibration level
Random vibration
Vibration frequency
Acceleration rms value
10 to 55 Hz
20.0 m/s
2
, 3 axis, 20 cycles each
5 to 500 Hz
X, Y axis 6.5 m/s
2
, Z axis 10.2 m/s
2
, 1 hour each
Backup battery life
Approx. 10 years (ambient temperature 23°C), for clock backup
Standards
Safety standards
EMC
EN61010-1
Overvoltage Category II (CAT II)
*
2
EN61010-2-30
Measurement Category
*
3
Dependent on specifications of
installed modules
EN61326-1, class A
Dimensions
Approx. 394 mm (W) x 334 mm (H) x 199 mm (D)
*
Excluding protrusions
Mass
9.5 kg or less (main unit only)
Warranty period
1 year
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...