9.Using Optional Modules - 9.3.Two Channel High-Speed Voltage Module (RA30-103)
1WMPD4004444
109
Input module
RA3100 main unit
Slots
1ch H
L
±500 V
AC, DC 300 V
±500 V
2ch H
L
Between each input and
the chassis, to ground
AC, DC 300 V
2 Input frequency and input voltage derating characteristics
A maximum of 1000 Vp-p can be input for the voltage and 5 MHz for the frequency, but there are
restrictions on the relationship between the voltage and frequency. These set limits to avoid damage to
devices and enable safe measurement.
Increased device heat and input voltage may damage the signal source and device.
3 Maximum rated voltage to ground
CAUTION
Use it within the range of safe usage.
Take care, as the input current will be approximately 21 mA when at 1000 Vp-p 100 kHz, which will
apply load on the signal source.
Ensure that the voltage between each input and the chassis (GND) and between each channel does
not exceed 300 V AC or DC. (Damage may be caused if the voltage is exceeded.)
0
200
400
600
800
1000
10
100
1000
10000
Input
m
ax
im
um
v
ol
tage
(Vp
-p)
Input frequency (kHz)
Frequency - voltage derating characteristics
損傷限界
安全使用範囲
Damage limit
Safe usage range