11.Maintenance -11.2.Backing Up Recorded Data
1WMPD4004444
139
11.2. Backing Up Recorded Data
This product records measured data on the internal SSD. Make sure to
periodically perform maintenance (data backup or deletion) on the SSD, because
failing to do so may prevent measurement from being performed due to
insufficient space.
The SSD remaining capacity monitor (SSD) is displayed on the side menu of the
monitor, and maintenance can be performed based on that display.
For information on backing up recorded data and deleting it from the SSD, see
"
8.2.1
<SSD remaining capacity monitor>
Displays the remaining capacity of the internal SSD in the indicator.
Maintenance procedure
Step 1. Copy the recorded data to back up to external media using the export function.
Step 2. Delete unnecessary recorded data.
11.2.1. Internal SSD Errors
The life of the internal SSD of the recording device greatly varies
according to the number of times data has been overwritten.
The health of the SSD can be checked in [Life remaining] and
[Health] in SSD on the [Maintenance] screen.
When the life remaining is close to 0%, please contact our sales
representative to replace the SSD.
11.3. Display Cleaning
When the surface of the display is dirty, wipe it clean with a soft, dry cloth or gauze dampened with ethanol.
11.4. Thermal Head Cleaning/Life
11.4.1. Cleaning
When recording has been performed for an extended period of time, material such as dust or printing
waste may adhere to the thermal head thermocouple. This may cause printing to become unclear and
reduce the image quality. In this case, follow the procedure below to clean the thermal head.
Step 1. Open the printer cover by pulling the lever of the printer block up.
Step 2. The thermal head can be seen in the top inside of the printer block. The thermocouple is at a line
4.4 mm from the thermal head edge. Clean that line area.
Step 3. Gently wipe it clean with a cotton bud or gauze dampened with ethanol. The recording paper colors
when exposed to ethanol, so it is recommended that you remove it from the stock area before cleaning.
Step 4. Load the recording paper after the thermal head
dries.
Lever
Thermal head (gold part)
Platen roller
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...