9.Using Optional Modules - 9.3.Two Channel High-Speed Voltage Module (RA30-103)
1WMPD4004444
107
(13) Disp. max: Set (by tapping the key and turning the knob) the display upper limit value (scale value) of
the top of the display range.
(14) Zero adjust: Cancels the input offset of the input channel. Execute zero cancellation to perform more
accurate measurement.
(15) Sheet: Set the monitor display/printer print sheet of the set channel.
(16) The available range and current measurement value monitor.
9.3.3. Measurement Setup
Check the slot number and channel, connect the signal to the corresponding input module, and enable
Meas. to display the input waveform on the monitor screen. Perform adjustment for the required
signal waveform while viewing the waveform.
Follow the procedure below to set the input channel.
Step 1. Set the input coupling in with the Coupling key (6).
Coupling
Contents
DC
Enables measurement of the actual input signal, including the DC and AC
component.
Set DC coupling when performing measurement.
AC
Measures the AC component of the input signal only. Set this when you want to
measure only the amplitude of an AC signal, as it cancels the DC offset of the signal.
GND
Connects the channel input to GND without connecting the input signal inside the
channel.
Enables the input GND level to be checked with waveform monitoring or printer
recording.
Step 2. Set the measurement range according to the target for measurement.
The input sensitivity can be changed with the Meas. range key (7) in the channel setup sub
menu.
The value displayed for the measurement range (RANGE) indicates the input (measurement)
maximum value and corresponds to 10 div on the waveform monitor. When the display position is
50%, ±RANGE (full measurement range) is displayed.
Tap the Meas. range key and turn the rotary knob to change the range. Turn the rotary knob
counterclockwise to reduce the sensitivity and clockwise to increase the sensitivity. The following 8
measurement ranges are available.
500 V 200 V 100 V 50 V 20 V 10 V
5 V
2 V
1 V 500 mV 200 mV 100 mV
When the input exceeds the measurement range, [Over range] is displayed on the bottom right of
the screen, and the main unit emits a warning beep. Reduce the sensitivity with the measurement
range so that the input signal does not exceed the range.
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...