9.Using Optional Modules - 9.3.Two Channel High-Speed Voltage Module (RA30-103)
106
1WMPD4004444
9.3.
Two Channel High-Speed Voltage Module (RA30-103)
9.3.1. Overview
This two channel voltage input module samples a DC to 5 MHz signal at 16-bit 20 MS/s within the
measurement range of ±100 mV to ±500 V and can perform A/D conversion. It includes an antialiasing
filter used for FFT processing and analog filter for waveform observation. It is insulated between each
channel and between input and output.
9.3.2. Setup the Input Channel
<This section describes the setting switches and setting values of the RA3100 main unit. >
(1) Slot number, input module type
(2) Change slot:
You can change the display slot by swiping this
sub menu left or right or tapping the left <
and right > key.
(3) Select channel:
Select the channel in the slot.
(4) Meas. ON/OFF
ON:
Measure and record the input signal.
(5) Color:
Change the display color of the waveform
monitor.
(6) Coupling: Switch the input signal coupling in the order DC
→ GND → AC. Tap the key and turn the rotary
knob to select.
(7) Meas. range:
Change the measurement range of the input
channel.
When this key is tapped, the rotary knob is
enabled (the LED lights up) and the range can
be selected by turning the knob.
(8) L.P.F.:
Change the low-pass filter of the input channel.
When this key is tapped, the rotary knob is
enabled (the LED lights up) and the filter can be
selected by turning the knob.
(1)
(2)
(2)
(3)
(3)
(4)
(5)
(6)
(7)
(8)
(10)
(11)
(12)
(13)
(14)
(15)
(16)
(10) Disp. pos.: Specify the position of the waveform monitor to display the specified range of the waveform
display area. Specified as a percentage indicating the center position of the display range
from the bottom of the monitor when the full range of the monitor is 100%.
(11) Disp. range: The display width in the amplitude direction of the waveform display area. Specified as the
percentage of the display width when the full range of the monitor is 100%.
Example When 50% is set, the waveform display is 10 div of the total width of 20 div.
(12) Disp. min: Set (by tapping the key and turning the knob) the display lower limit value (scale value) of
the bottom of the display range.
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...