2.Pre-Measurement Procedures - 2.2.Turning the Power On/Off
32
1WMPD4004444
2.2.
Turning the Power On/Off
2.2.1. Connecting the AC Power Cable
Be sure to check the following points before connecting the AC power cable to this product.
Make sure that the power supply matches the rating indicated on the rating plate attached to this
product.
Ensure amp or interface units are inserted securely.
WARNING
This product must be grounded before turning on the power.
This grounding protection is for the safety of this product, as well as for that of the user and peripheral
equipment.
If the AC power cable that comes with this product is connected to a 3-pin power outlet equipped with
a protective conductor pin, the product is automatically grounded.
Do not use an extension cable without protective grounding.
Do not use this product when protective grounding cannot be performed using a power outlet that
matches the supplied AC power cable.
NOTE
AC power cable
The AC power cable that is included in this product (1KO6165-200: 125-V AC system, 2 m) is a 3-pin
type which has a round pin at the center for protective grounding.
2.2.2. Turning On the Power
Step 1. Perform the check indicated below.
<Items to check before turning on the power>
Is this product installed in a safe location?
Is the usage environment OK?
Is the top of the touch panel free from any writing implements or tools, etc.?
Step 2. Upon confirming that all of the above check items are fine, connect the inlet side of the AC power
cable to the AC socket of this product.
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...