![background image](http://html1.mh-extra.com/html/aandd/ra3100-omniace/ra3100-omniace_instruction-manual_3072739064.webp)
6.Measuring Input Signals - 6.3.Starting and Ending Recording
64
1WMPD4004444
(7) Start time + Interval (N times)
Interval time
Press START key
Start time
1 time
N times End recording
Recording time
Recording time
During record
During record
Printer recording
SSD recording
Memory recording
Trigger occurred Trigger occurred Trigger occurred
When the START key on the operation panel is pressed, the product enters the measurement
standby state. Recording starts at the time set in Start time , then continues recording for the time set
in Recording time .
Measurement repeats at the interval set in Interval time .
(8) Start time + Start t Interval (N times)
Start trigger occurs
Interval time
Start trigger occurs
Press START key
1 time
N times End recording
Recording time
Recording time
Start time
During record
During record
Printer recording
SSD recording
Memory recording
Trigger occurred Trigger occurred Trigger occurred
When the START key on the operation panel is pressed, the product enters the measurement
standby state and starts the procedure indicated in "(6) Start t Interval (N times)" at the time set
in Start time .