9.Using Optional Modules - 9.6.Remote Control Module (RA30-112)
1WMPD4004444
123
9.6.3. Setting the Input Channel
Tap the CH key on the side menu to display the channel settings
menu. "SLOT 9" where the remote module (RA30-112) is installed in
slot selection is displayed.
(1) The slot number of SLOT 9.
(2) Response:
Select [High speed], [Medium speed], or [Low
speed] for the response speed of the remote
input signal.
(3) OSC: Select the excitation signal used for each
module when the AC strain module is installed.
(4) TRIG: Sets the TRIG signal for remote input.
(1)
(2)
(3)
(4)
9.6.4. Standard Clock Output
A standard clock is output from the output terminal of this module at 0 to 5 V and 1 kHz.
This clock signal can be connected to the voltage input module and monitored to check module
operation.
9.6.5. Measurement Setup
1 Response
Select [High speed], [Medium speed], or [Low speed] for the response speed of the remote input signal.
In noisy environments, select [Low speed] to perform control with a filter.
The effective pulse width of the input signal is as follows.
For high speed response:
High period
1 μs or more,
low period
1 μs or more
For medium speed response: High period 1 ms or more, low period 1 ms or more
For low speed response:
High period 10 ms or more, low period 10 ms or more
2 OSC
Select the excitation signal used for the RA3100 main unit when the AC strain module is installed.
Internal:
Use the OSC signal generated in the RA3100 main unit where this module is installed.
It can also be supplied to another RA3100 connected with synchronization.
External: Sends the ACCLK IN signal input from remote input to the implementation module.
See the connection diagram in "
3 TRIG
Sets the TRIG signal for remote input.
OFF:
Do not use the TRIG signal for remote input.
Start trigger:
Input the TRIG signal for remote input as the start trigger.
Memory trigger: Input the TRIG signal for remote input as the memory trigger.
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...