10. 10Appendix - 10.1. Sampling Data Format
128
1WMPD4004444
T0
T1
T2
T3
T4
T5
T6
T7
T8
T9
Analog input waveform
Sampling period
Time
T0
T1
T2
T3
T4
T5
T6
T7
T8
T9
Sampling data
Sampling period
Time
10. Appendix
This chapter provides additional explanations of various functions.
10.1. Sampling Data Format
This product has two data formats: normal sampling and P-P sampling.
10.1.1. Normal Sampling
With normal sampling, the A/D value of the sampling period is recorded as data and used for waveform
reproduction and data analysis.
Playback the input waveform from sampling data
If the input signal is too fast for the sampling period, the waveform reproducibility may drop and lead to
the unexpected loss of pulses.
Raising the sampling speed improves the waveform reproducibility but increases the recorded data.
10.1.2. P-P Sampling
P-P sampling performs sampling with the data within the sampling period as the fastest sampling (20
MS/s), to detect the peak values (maximum value and minimum value) and record those peak values as
the recorded data.
This data format is suitable for the waveform playback of long-term recording, as it enables waveform
reproducibility of a wide band of data, without sudden spike noise, etc.
A disadvantage is that it cannot perform data analysis after recording (average, RMS, FFT, etc.)
Sampling period
Sampling data
T0 T2
T4 T6
T8
T10 T12 T14 T16 T18
T0
T1
T2
T3
T4
T5
T6
T7
T8
T9
Time
Sampling period
P-P data
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...