9.Using Optional Modules - 9.5.Two Channel Temperature Module (RA30-106)
1WMPD4004444
121
Input module
RA3100 main unit
Slots
1ch +/A
-/B
S/b
AC, DC 300 V
2ch +/A
-/B
S/b
Between each input and
the chassis, to ground
AC, DC 300 V
GND (three pin power cable connection)
6 LFD function
This module has an LFD function that can be enabled or disabled when performing TC measurement.
When enabled, 0.5 μA current is supplied and + side over range output occurs when the sensor is open
(at 300 kΩ or higher). (Over range detects the possibility
of a disconnection.)
NOTE
When LFD is enabled, a temperature rise error will occur from the power consumption in the
thermocouple (TC) resistance because a current of 0.5 μA is constantly supplied.
It can be disabled when performing standard measurement to reduce the above error.
When the setting is enabled, detection is possible when the open resistance is 300 kΩ or higher.
7 Maximum rated voltage to ground
9.5.5. Spare Parts
Two sets of temperature sensor connectors are included as standard, and they can be connected to an
alternate sensor in advance to allow easy changing.
Name (type)
Shape/characteristics
Remarks
Temperature sensor
connector
Top cable cover
Bottom cable cover
Two sets of cable ties
RA30-555
Can be
connected/disconn
ected to/from the
input connectors of
the connector
module for
temperature sensor
connection
RA30-106
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...