5.Trigger Setup - 5.2.Memory Trigger
1WMPD4004444
45
Step 3. Up to 18 trigger sources ( T1 to T18 ) can be set.
Tap the number of the trigger source to set to display the details screen.
(1)
(2)
(3)
(4)
(5)
(6)
(1) Trigger menu selection:
Switches between the Memory trigger, Start trigger, or Memory block menu
(2) Channel: Selects the TRIGn source channel.
(3) Detection: Selects UP, DOWN, INTO WIN, or OUT WIN for the polarity of the trigger signal.
UP
The trigger is detected when the value exceeds the trigger level (threshold).
DOWN
The trigger is detected when the value is below the trigger level (threshold).
INTO WIN The trigger is detected when the value enters the range of the upper limit
value or lower limit value of the trigger level.
OUT WIN The trigger is detected when the value leaves the range of the upper limit
value or lower limit value of the trigger level.
(4) Threshold: Sets the trigger level (threshold).
For INTO WIN / OUT WIN, there are two settings: the upper threshold and lower threshold.
(5) Filter:
Sets the filter time for noise removal.
(6) Close:
Ends the setting operation and returns to the trigger list.
CAUTION
The trigger level is a value relative to the set measurement range. The value also changes when the
measurement range is changed.
(Example) When the trigger level is set to 10 mV when the range is 100 mV, and then the
measurement range is changed to 200 mV, the trigger level is changed to 20 mV.
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...