11.Maintenance - 11.1.Managing/Handling Recording Paper and Printer Recorded Data
1WMPD4004444
137
11. Maintenance
The frame must not be removed from this product other than by our service engineers, as the product is
a precision device.
This section describes the maintenance of the product.
11.1. Managing/Handling Recording Paper and Printer
Recorded Data
11.1.1. Replacing Recording Paper and Monitoring Remaining Paper
Use the chart recording paper specified by A&D. Use of a chart that is not recommended may cause
failure in printing or shorten the life of the thermal head.
Red is printed on the recording paper when the recording paper is running low. The remaining paper
monitor (PAPER) is displayed on the side menu of the monitor. Follow these to replace the recording
paper.
For information on the replacement method, see "2.1.3.
".
Remaining Paper Monitor Operations
<After replacing recording paper>
<During use>
<No paper remaining>
Reset the remaining paper monitor after replacing the recording paper. Press and hold PAPER
(remaining paper monitor) on the side menu to display the dialog box for resetting the remaining paper
monitor.
Tap Reset to 30 m for a 30 m paper roll (YPS-106 or YPS-108)
or Reset to 200 m for a 200 m paper roll (YPS-112), and then tap CLOSE .
Reset to 30 m
Reset to 200 m