9.Using Optional Modules - 9.5.Two Channel Temperature Module (RA30-106)
120
1WMPD4004444
Step 2. Insert the top cable cover and bottom cable cover in the socket.
Secure the cable from the bottom cable cover with a cable tie.
Top cable cover (included)
Temperature sensor connectors
Bottom cable cover (included)
Secure with cable tie (included)
Step 3. Connect the socket to the connector of the input panel.
Step 4. Secure the socket to the connector of the input panel using the socket fixing screw.
Input connectors
Connector fixing screw
NOTE
Socket fixing screw tightening torque: 0.2 Nm to 0.25 Nm
The cable cover protects the terminal and cable fixing screw areas from static electricity, etc.
5 Reference junction compensation (CJC) when measuring thermocouple (TC)
Reference junction compensation is required when measuring the thermocouple (TC), and can be
switched between internal and external compensation with this module.
Internal reference junction compensation measures the reference junction temperature at the module
front panel.
When external compensation is set, it is necessary to externally perform 0°C reference junction
compensation.
NOTE
Perform measurement so that the temperature around the input connectors is stable.
If there is a rapid change in the surrounding temperature, wait until the module temperature stabilizes
(about one hour) and then start measurement.
Ensure that the input connectors are not directly exposed to wind.
Do not block the air holes of the front panel.
When performing external reference junction compensation
Compatible extension wire
Thermocouple
Temperature measurement point
0°C
External 0°C reference junction
+
-
Summary of Contents for RA3100 Omniace
Page 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Page 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Page 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Page 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...