2 • Features
Doc. 031857-03 4/06
2-37
details, refer to the EluGen cartridge manual. Cartridge manuals are
included on the Dionex Reference Library CD-ROM (P/N 053891).
EluGen Cartridge Holder
The EluGen cartridge is installed in a cartridge holder, which mounts on
the ICS-2000 top cover (see
). The cartridge holder also houses
the high-pressure degas tubing assembly.
The ICS-2000 can control up to two EluGen cartridges. However, when
operating with an EGC II K
2
CO
3
EluGen Cartridge and the pH Modifier,
EluGen
Cartridges
Eluent Concentration Range
Comment
K
2
CO
3
/EPM
Electrolytic
pH Modifier
0.1 to 15 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 2.0 mL/min
flow
X = 15/flow
The total of the eluent
concentrations from
both the EluGen
cartridge and the EPM
must not exceed the
specified range. The
EPM concentration
must not exceed
10 mM. See
“Generating
Carbonate/Bicarbonat
e Eluent” on page 2-
39
for details.
KOH/KOH
KOH/MSA
KOH/NaOH
MSA/MSA
MSA/NaOH
NaOH/NaOH
0.1 to 50 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 3.0 mL/min
flow
X = 50/flow
The eluent
concentration range
for each cartridge is
cut by 50%.
LiOH/LiOH
0.1 to 40 mM at 0.1 to 1.0 mL/min
flow
0.1 to X mM at 1.0 to 3.0 mL/min
flow
X = 40/flow
The eluent
concentration range
for each cartridge is
cut by 50%.
Table 2-2. Eluent Concentration Ranges for Linked
Dual-Cartridge Configurations
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...