ICS-2000 Ion Chromatography System
B-108
Doc. 031857-03 4/06
NOTE The ICS-2000 and the AS or AS50 (USB) can be set up
for stand-alone (front-panel) operation. This requires a
TTL cable connection between the AS or AS50 (USB)
and the ICS-2000. See
for stand-alone setup
instructions.
B.18.1 AS or AS50 (USB) Connections
1. Place the autosampler to the left of the ICS-2000 on the workbench.
2. Open the ICS-2000 and autosampler doors.
3. The needle seal line that extends from the inject port in the
autosampler is coiled inside the autosampler door, on the right side.
Route this line through the autosampler right side slot (see
) and the ICS-2000 left side slot to the ICS-2000
injection valve.
The needle seal line from the autosampler inject port to the ICS-2000
injection valve is calibrated for use with the ICS-2000. To ensure
injection precision, do not shorten (by cutting) or lengthen this line, or
replace it with a different (uncalibrated) line.
Figure B-104. Autosampler Drip Tray