ICS-2000 Ion Chromatography System
4-24
Doc. 031857-03 4/06
3. Under
Degas Mode
,
verify that the
Always Off
option is
not
selected. If
it is, select one of the following settings:
•
Cycle
, and select
On
: 30 seconds and
Off
: 10 minutes
•
Monitor
4. If the
Degas Mode
settings are correct, but the degas assembly still does
not run, test the assembly by selecting the
Always On
option. The pump
should turn on immediately. If it does not, the vacuum degas assembly
may need to be replaced. Contact Dionex for assistance.
NOTE Degas Mode options can also be selected from the touch
screen
PUMP
page (see
5. If the degas pump runs when you select
Always On
, but does not run
when
Monitor
mode is selected, recalibrate the degas assembly (see
Select the Degas Mode, Always On option only when testing the vacuum
degas assembly. Do not select the Always On option for routine use.