5 • Service
Doc. 031857-03 4/06
5-59
Conditioning the EluGen Cartridge
Always condition the EluGen cartridge for 30 minutes before initial use.
1. Make sure the eluent reservoir is filled with ASTM filtered, Type I
(18-megohm) deionized water.
2. Verify that the suppressor is off.
3. Verify that the
Automatically turn on with pump
option is not
selected on the ICS-2000 touch screen
SUPPRESSOR
). Disabling this option lets you turn on the pump, without
automatically turning on the suppressor.
4. From the ICS-2000 Control panel in Chromeleon or the ICS-2000
touch screen
HOME
page, set the pump flow rate to
1.00 mL/min
.
5. Turn on the pump.
6. Select the following settings from the Chromeleon Control panel or
the touch screen
ELUENT GENERATOR
page:
•
Set the
EGC-1
concentration to
9 mM
.
•
Set the
EGC-2
concentration to
1 mM
.
7. Turn on the EGC-1 and EGC-2 currents.
8. Run at these settings for 30 minutes.
9. Turn off the EGC-1 and EGC-2 currents.
10. Turn off the pump.
11. Remove the backpressure coil.
12. Reconnect the
TO INJ VALVE IN - P/ELUENT OUT
line to the mixer
inlet.
13. Reconnect the
TO SRS/AES REGEN OUT
line to the suppressor
REGEN
OUT
port.
To avoid damaging the suppressor, always turn off the suppressor
before conditioning the EluGen cartridge. The pump flow is on during
conditioning, however, no flow reaches the suppressor.
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...