ICS-2000 Ion Chromatography System
5-34
Doc. 031857-03 4/06
6. Pull the valve straight out
from the component panel and
begin pulling the attached
cable out of the opening in the
panel.
7. The cable is connected to a
matching cable inside the ICS-
2000. Continue pulling until
the connectors for the two
cables are outside the panel
(see
). Disconnect
the cables.
Figure 5-25. Eluent Valve Cable
8. Remove the mounting plate on
the old valve by unscrewing
and removing the two screws.
Save the screws.
9. Thread the cable from the new
eluent valve through the
mounting plate.
10. Attach the mounting plate to
the new valve using the
screws removed in
.
Figure 5-26. Removing the Valve Mounting Plate
11. Connect the cable from the new valve (P/N 057945) to the ICS-2000 cable.
Feed the cables back inside the ICS-2000.
12. Align the new valve on the component panel with the liquid ports facing up
and attach it with the two screws removed in
. Reconnect the liquid
lines.
13. Turn on the power to the ICS-2000.
14. Prime the pump (see
Disconnect
Cables
Eluent Valve
Cable
Remove
Screws
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...