Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-77
1. Connect the
TO INJ VALVE IN - P/ELUENT OUT
line to the coupler
attached to the mixer inlet line. Route this line along the back of the
drip tray to prevent it from being pinched when the ICS-2000 door is
opened and closed
2. Route the mixer outlet line to the back of the component panel (see
3. Connect the mixer outlet line to port
P
(
2
) on the injection valve.
NOTE Backpressure tubing may be required between the
injection valve and the EGC CO
3
Mixer outlet. The
EluGen cartridge requires 14 to 16 MPa (2000 to
2300 psi) of system backpressure for removal of
electrolysis gas from the eluent produced by the
cartridge. See
for details.
Go on to
Figure B-75. Mixer Outlet Routed to the Back of the Component Panel
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...