Connecting the ICS-2000 to the Chromeleon PC
Doc. 031857-03 4/06
B-11
B.5.2
Connecting Additional USB Devices
The ICS-2000 provides two USB ports for connecting additional Dionex
USB-compliant modules to the system.
To connect an additional USB module, plug the “A” connector on a USB
cable into an “A” port (labeled
TO MODULE
) on the ICS-2000 rear panel
and plug the cable’s “B” connector into the USB receptacle on the USB
module (see
NOTE If the number of USB devices in the system exceeds the
number of available USB ports, connect the devices to
the system with a USB-compliant, full-speed external
hub. One or more hubs is also required if the ICS-2000
or other USB device is more than 5 meters (16 ft) from
the PC. Refer to the manual provided by the hub vendor
for complete installation instructions.
Figure B-10. Example Connections: Two Dionex USB Modules
Connected to the ICS-2000
The USB standard limits the USB cable length to 5 meters (16 ft). Each
USB device can be separated from the PC by no more than five hubs.
Thus, each USB device can be located no more than 30 meters
(32 yds) from the PC.