Setting Up the Eluent Generator
Doc. 031857-03 4/06
B-39
illustrates the liquid flow path through the eluent generator
components after all setup steps are complete and the eluent generator is
ready for operation.
Figure B-31. Eluent Generator Flow Schematic with CR-TC:
Use for Generating KOH, LiOH, MSA, or NaOH Eluent
El
ue
n
t In
Re
ge
n
In
Re
ge
n
Out
Eluent
Out
CR-TC
EGC
DEGAS
EGC
IN
Connect to
Gas Separator
Assembly
TO DEGAS
ELUENT OUT
TO CR-TC
REGEN OUT
(blue)
TO SRS/AES
REGEN OUT
TO
C
R-T
C
RE
GE
N I
N
(o
ran
ge
)
WASTE,
GAS
SEPARATOR
DEGAS
ELUENT OUT
TO INJ
VALVE IN - P/
ELUENT OUT
Connect to
Inject Valve
Port P (2)
Connect to
Suppressor
Regen Out
Pre-plumbed
Pump Pulse
Damper
to
EGC II
KOH, LiOH,
MSA, or NaOH
Inlet
Outlet
EGC VENT
Send to waste
TO DEGAS
ELUENT IN
EGC
OUT
TO CR-TC
ELUENT IN
(red)
TO CR-TC/EPM
ELUENT OUT
(yellow)
TO PUMP/
DAMPER
*Add backpressure coils, if
needed, to create a system
backpressure of between 14 and
16 MPa (2000 and 2300 psi).
Back-
pressure
Coil*
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...