ICS-2000 Ion Chromatography System
2-32
Doc. 031857-03 4/06
2.4.2
Eluent Valve
2.4.3
Pump
The ICS-2000 pump is a microprocessor-based isocratic eluent delivery
system. Its variable speed, dual-piston series design ensures pulse-free
pumping for the most demanding applications.
Primary Pump Head
The primary pump head pumps eluent into the secondary head (see
). The check valves, which prevent reverse flow through the
pump, are located on the bottom (inlet) and top (outlet) of the primary
pump head. The priming valve is on the front of the pump head.
The eluent valve controls the flow
from the eluent reservoir. The
valve opens automatically when
the pump is started and closes
when the pump is turned off. The
valve can also be opened and
closed manually either from the
Chromeleon Control panel (see
), or the touch screen
PUMP
page (see
lets you perform service
procedures on pump components
without eluent leaks occurring.
Figure 2-17. Eluent Valve