D • TTL and Relay Control
Doc. 031857-03 4/06
D-9
Assign ICS-2000 Input Functions
Go to the ICS-2000
INPUT/OUTPUT
touch screen page and verify that the
Inject
function is assigned to TTL input 1, that Autozero is assigned to TTL input 2, and
that the
TTL Input Control Type
is
Normal Edge
Enable AS or AS50 (USB) Control of the Injection Valve
1. On the AS or AS50 (USB) front panel, press
Menu
and
0
to go to the
TIME
FUNCTION OUT
screen. In the
LOAD/INJECT VALVE: TTL 1
field, select
ENABLED
.
2. Turn the AS or AS50 (USB) power off and then back on.
Figure D-8. Input/Output Page
Figure D-9. AS or AS50 (USB) Time Function Out Screen