ICS-2000 Ion Chromatography System
B-90
Doc. 031857-03 4/06
1. Use a Phillips head screwdriver to
remove the existing small diameter
mounting clips from the inside of
the ICS-2000 door.
2. Install the larger diameter mounting
clips (P/N 063642) (see
Figure B-88. Location of Mounting Clips for the EGC CO3 Mixer
3. Orient the mixer with the outlet
pointing upward and push it onto the
clips (see
outlet must point upward to ensure
thorough mixing of the eluent.
4. Locate the 35 cm (14 in) length of
black PEEK tubing (white labels)
(P/N 064346). Connect the end
labeled
TO
EGC CO3 MIXER IN
to the
mixer inlet.
5. Locate the 60 cm (24 in) length of
black PEEK tubing (white labels)
(P/N 064343). Connect the end
labeled
TO EGC CO3 MIXER OUT
to
the mixer outlet.
Figure B-89. EGC CO
3
Mixer Installed on the ICS-2000 Door
Mixer
Mounting
Clips
Mixer Inlet
Mixer Outlet