4 • Troubleshooting
Doc. 031857-03 4/06
4-11
This error occurs if you turn on the suppressor but the ICS-2000 cannot
establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection (see
2. If the error persists, replace the suppressor (see
This error occurs when, in order to maintain the selected current, the ICS-
2000 is required to apply a higher voltage than the suppressor can support.
To troubleshoot:
•
Reduce the flow rate.
•
Rehydrate the suppressor. Refer to the suppressor manual for instructions.
Suppressor manuals are included on the Dionex Reference Library CD-
ROM (P/N 053891).
•
If the error persists, replace the suppressor (see
This error occurs if you turn on the suppressor and the ICS-2000 cannot
establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection (see
2. If the error persists, contact Dionex for assistance. There may be a
problem in the suppressor controller electronics. The ICS-2000
electronics components cannot be serviced by the user.
SUPPRESSOR OVER VOLTAGE
SUPPRESSOR OVER POWER
SUPPRESSOR NOT CONNECTED