ICS-2000 Ion Chromatography System
B-120
Doc. 031857-03 4/06
B.24 Installation Troubleshooting
Problem:
When the ICS-2000 power is turned on for the first time, a Windows
message box appears asking for a USB configuration file (cmwdmusb.inf).
Possible Cause:
The USB cable was connected and the power turned on before
Chromeleon was installed.
Solutions:
Follow these steps to resolve the problem:
1. Click
Cancel
in the Windows message box.
2. Turn off the power to the USB device and unplug the USB cable from the
computer.
3. Install Chromeleon (see
4. Reconnect the USB cable to the computer and turn on the power to the
device. Windows will now automatically recognize the USB device.