ICS-2000 Ion Chromatography System
3-6
Doc. 031857-03 4/06
shows a connected Control panel for an ICS-2000 system.
Figure 3-6. ICS-2000 System Control Panel (Connected)
3.4
Set Up the Eluent Reservoir
3.4.1
Filter the Deionized Water
Filtering removes small particulates in the deionized water that may
contaminate the pump check valves and cause erratic flow rates or loss of
prime. An end-line filter (P/N 045987) is provided in the ICS-2000 Ship
Kit (P/N 064375) for this purpose.
Install the end-line filter on the end of the deionized water line, inside the
reservoir. Verify that the end of the filter extends to the bottom of the
reservoir and that the filter is submerged in deionized water. This prevents
air from being drawn through the lines.