5 • Service
Doc. 031857-03 4/06
5-65
5.19.1 Recording the EPM Serial Number in Chromeleon
The serial number is printed on a label on the EPM.
1. Open the Server Configuration program (select
Start > Programs
(or
All Programs
)
> Chromeleon > Server Configuration
).
2. Double-click the ICS-2000 system icon. The properties dialog box for
the ICS-2000 appears.
3. Click the
Eluent Generator
4. Select the
EGC-2
check box.
5. Enter the serial number of the EPM.
6. Verify that the cartridge type is
pH Mod
. If not, make sure the serial
number was entered correctly.
7. Click
OK.
Figure 5-59. Eluent Generator Configuration Properties Example