ICS-2000 Ion Chromatography System
B-62
Doc. 031857-03 4/06
4. Screw the luer adapter
onto the vent opening.
Remove the cap on
the luer adapter by
twisting.
Save the
cap.
5. Twist the luer lock
and EluGen cartridge
vent line into the luer
adapter (see
Figure B-60. Installing the EluGen Cartridge Vent Line
NOTE To prevent leaks from the EluGen cartridge,
always replace the luer adapter cap before
moving the EluGen cartridge.
6. Direct the vent line to the waste container and insert it down into the
gas separator waste tube about 10 to 15 cm (4 to 6 in).
Installing the EPM
1. Remove the plugs from
the EPM eluent and
regen liquid ports.
2. Push the EPM onto the
ball studs on the
mounting bracket (see
Figure B-61. EPM and Mounting Bracket
Luer Adapter
EGC Vent Line
Luer Lock
EGC Cable
Ball Studs
Mounting Bracket
EPM