ICS-2000 Ion Chromatography System
5-88
Doc. 031857-03 4/06
•
Connect the line labeled
TO INJ VALVE IN - P/ELUENT OUT
to the
EGC CO
3
Mixer inlet.
•
Connect the line labeled
TO PUMP/DAMPER
to the pulse damper.
4.
Direct the
WASTE, GAS SEPARATOR
line from the top of the ICS-
2000 to the rear panel. Snap the line onto one of the tubing clips on
the rear panel. Connect the line to the gas separator waste tube (see
5.
Go on to
5.21.5 Reinstalling the EluGen Cartridge
1. Pull the coiled black
EGC IN
and
EGC OUT
lines out from inside the
EGC holder. Disconnect the
EGC IN
line from the coupler that
connects it to the
EGC OUT
line.
2. Connect the
EGC IN
line
to the EluGen
cartridge
INLET
port
3. Remove the coupler
and connect the
EGC
OUT
line to the
EluGen cartridge
OUTLET
port.
Figure 5-77. EluGen Cartridge Inlet and Outlet Port Connections
4. Turn over the EluGen cartridge (fittings facing down). Check for any
bubbles in the eluent generation chamber, and if necessary, shake and
tap the EluGen cartridge to remove them.
5. Reinstall the EluGen cartridge in the holder.
6. Reconnect the EluGen cartridge cable and the CR-TC or EPM cable
(if installed).
7. Remove the cap from the luer adapter on the EluGen cartridge vent
opening and re-connect the
EGC VENT
line.
EGC OUT
OUTLET
Port
EGC IN
INLET
Port
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...