C • Touch Screen Operation
Doc. 031857-03 4/06
C-19
C.8
Eluent Generator Page
Use the
ELUENT GENERATOR
page to set the eluent concentration of each
installed EluGen cartridge (EGC). You can also enter the serial number and view
the expiration date and remaining life of each cartridge. This page also provides
on/off control of a Continuously Regenerated Trap Column (CR-TC). To open the
page, touch the
EGC
button on the
HOME
page or select
EGC
from the menu of
pages.
C.8.1
EGC Serial Number
The
Serial Number
field displays the serial number of the EluGen
cartridge currently installed. After installing a new EluGen cartridge,
enter the cartridge serial number. If you use Chromeleon to control the
ICS-2000, enter the serial number in the Chromeleon Server
Configuration (see
), not on the
ELUENT GENERATOR
page. If
you enter the serial number first in Chromeleon, the number is displayed
on the
ELUENT GENERATOR
page. However, if you enter the serial
number on the
ELUENT GENERATOR
page, it will not be available in
Chromeleon.
Figure C-16. Eluent Generator Page
Touch here
to enter a
new serial
number
Touch here
to drop-
down a list
of existing
serial
numbers
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...