ICS-2000 Ion Chromatography System
B-24
Doc. 031857-03 4/06
4. Unscrew the six
thumbscrews on the
top metal plate (see
remove the plate.
5. Go on to
.
Figure B-21. ICS-2000 Column Heater Cover Removed
B.9.2
Installing the Columns
1. Remove the coupler connecting
GUARD IN
and
GUARD OUT
.
2. Connect
GUARD IN
to the guard column inlet. Check the arrow on the
column label; it should point down (flow away from the injection
valve).
3. Connect
GUARD OUT
to the guard column outlet.
4. Push the guard column onto the back column clip.
5. Remove the coupler connecting
COL IN
and
COL OUT
.
6. Connect
COL IN
to the separator column inlet. Connect
COL OUT
to
the separator column outlet. Check the arrow on the column label. It
should point up (flow toward the cell).
7. Push the separator column onto the front column clips.
8. Replace the top metal plate on the column heater.
Thumbscrew (6)