5 • Service
Doc. 031857-03 4/06
5-73
It takes about 5 minutes to fill a 4-mm mixer and about 2 minutes to
fill a 2-mm mixer.
12. Turn off the pump.
5.20.3 Filling the EGC CO
3
Mixer with Eluent
Before initial use, follow the steps in this section to fill the EGC CO
3
Mixer with the eluent required for the application to be run. The
procedure varies slightly, depending on the type of eluent to be generated:
•
For carbonate eluent, see
“Filling the EGC CO3 Mixer with
below.
•
For carbonate/bicarbonate eluent, see
with Carbonate/Bicarbonate Eluent” on page 5-77
.