5 • Service
Doc. 031857-03 4/06
5-47
11. Lift the EluGen
cartridge out of the
holder, turn it over,
and set it in the
service area.
12. Disconnect the
backpressure coil
from the EluGen
cartridge
OUTLET
port
.
13. Connect the
EGC
OUT
line from the
holder to the
OUTLET
Figure 5-41. EluGen Cartridge Inlet and Outlet Connections
14. Turn over the EluGen cartridge (fittings facing down). Check for
bubbles. If necessary, shake and tap the EluGen cartridge to remove
bubbles.
15. Reinstall the EluGen cartridge in the holder.
16. Reconnect the EluGen cartridge cable.
17. Reconnect the CR-TC cable.
18. If you disabled the
Automatically turn on with pump
option in
, select the option.
EGC OUT
OUTLET
Port
EGC IN
INLET
Port