ICS-2000 Ion Chromatography System
B-60
Doc. 031857-03 4/06
11. Turn the EluGen cartridge over (with the fittings facing down).
Shake
the EluGen cartridge vigorously, and tap it with the palm of your
hand 10 to 15 times. Check that all bubbles trapped in the eluent
generation chamber are dislodged.
12. Slide the EluGen cartridge down into the
holder until it stops (see
Make sure the cable remains out of the
holder.
NOTE To avoid crimping the lines, do not
twist the cartridge when you slide it
into the holder. The lines are
designed to coil inside the holder.
Figure B-57. EluGen Cartridge Installed in the Holder
The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), a corrosive acid (MSA), or a concentrated
K
2
CO
3
solution. Wear protective eyewear and gloves when handling
the cartridge.
La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), un acide corrosif (MSA), ou une solution
concentrée de K
2
CO
3
. Porter des lunettes et des gants protectives en
manipulant la cartouche.
Die EluGen-Kassette enthält eine korrodierende Base (KOH, LiOH
oder NaOH), eine korrodierende Säure (MSA) oder eine konzentrierte
K
2
CO
3
-Lösung. Tragen Sie daher beim Umgang mit Kassette eine
Schutzbrille und Handschuhe.
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...