4 • Troubleshooting
Doc. 031857-03 4/06
4-21
•
No flow from pump
This condition has several possible causes. See
and
for details.
•
Cell electronics malfunctioning
Use the Chromeleon Wellness Panel diagnostics to test the electronics (see
and the Chromeleon online Help for instructions). If the variance
reading is outside the tolerance range (less than 1
µ
S), the electronics are
malfunctioning. Contact Dionex for assistance. The ICS-2000 electronics
cannot be serviced by the user.
4.12 High Cell Output
•
Background not suppressed by suppressor
Verify that the suppressor is turned on and the current is set to the correct
value. Refer to the suppressor manual for additional troubleshooting
guidance. Suppressor manuals are included on the Dionex Reference Library
CD-ROM (P/N 053891).
•
Sample concentration too high
Dilute the sample
•
Wrong eluent
Check that you are using the correct eluent for your application.
•
Background conductivity not offset from conductivity reading
Before making an injection, allow the background conductivity to equilibrate,
and then press
Autozero
on the Chromeleon Control panel or the ICS-2000
touch screen
HOME
page.
•
Cell out of calibration
Recalibrate the cell (see
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...