4 • Troubleshooting
Doc. 031857-03 4/06
4-17
•
Flow rate is set to 0
Set the flow rate from either the Chromeleon Control panel or the ICS-2000
touch screen
HOME
page.
4.5
No Flow
•
Pump waste or priming valve open (see
Close the valves by turning the knobs clockwise until fingertight.
DO NOT
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.
•
Flow rate is set to 0
Set the flow rate from the Chromeleon Control panel or from the touch screen
HOME
page.
•
Eluent valve is closed
Open the eluent valve from the Chromeleon Control panel or from the touch
screen
PUMP
page.
•
Pump not primed
•
Broken pump piston
Replace the piston (see
Figure 4-2. Waste and Priming Valves
Waste Valve
Priming Valve