ICS-2000 Ion Chromatography System
B-74
Doc. 031857-03 4/06
1. Disconnect the pump outlet from the mixer inlet.
2. Reconnect the pump outlet and the EluGen cartridge inlet.
3. Connect a 10-32 to 10-32 coupler to the mixer inlet.
4. Disconnect the outlet of the yellow PEEK backpressure coil from the
TO SRS/AES REGEN OUT
line and then connect the backpressure coil
5. Connect the mixer outlet to the
TO SRS/AES REGEN OUT
line.
NOTE The backpressure coil connection is temporary;
it is required only during initial filling of the
mixer.
6. Verify that the suppressor is off.
Figure B-73. Mixer Connected Directly to Eluent
Generator Outlet for Initial Filling
Back-
pressure
Coil
TO INJ PORT IN - P/ELUENT OUT
Mixer Inlet
Mixer Outlet
TO SRS/AES
REGEN OUT