ICS-2000 Ion Chromatography System
5-78
Doc. 031857-03 4/06
1. Locate the following items in the ICS-2000 Ship Kit (P/N 064375): a
yellow 0.5 mL/min, 7 MPa (1000 psi) backpressure coil
(P/N 053765), a 10-32 to 10-32 coupler (P/N 042627), and a 10-32 to
1/4-28 coupler (P/N 042806).
2. Use the 10-32 to 10-32 coupler to connect one end of the
backpressure coil to the
TO INJ VALVE IN - P/ELUENT OUT
line (see
3. Connect the other end of the backpressure coil to the coupler on the
mixer inlet.
4. Use the 10-32 to 1/4-28 coupler to connect the mixer outlet to the
TO
SRS/AES REGEN OUT
line.
NOTE The backpressure coil connection is temporary;
it is used only during initial filling of the mixer.
Figure 5-67. Mixer Connected Directly to Eluent
Generator Outlet for Initial Filling
Back-
pressure
Coil
TO INJ PORT IN - P/ELUENT OUT
Mixer Inlet
Mixer Outlet
TO SRS/AES
REGEN OUT