ICS-2000 Ion Chromatography System
B-110
Doc. 031857-03 4/06
B.18.2 AS or AS50 (USB) Chromeleon Setup
1. If you have not already done so, add the autosampler to the
Chromeleon timebase (see
2. In the ICS-2000 Server Configuration properties, verify that the
injection valve is controlled by the autosampler. To do this:
a.
Open the Chromeleon Server Configuration program and select
the ICS-2000 device in the timebase.
b. Select
Properties
on the
Edit
menu.
c.
Select the
Inject Valve
tab.
d. Verify that the
Pump_InjectValve
check box is selected and that
AS
is listed in the
Controlled By
e.
If
ICS-2000
is listed in the
Controlled By
column, select the
Pump_InjectValve
name and press the
F2
key. A dialog box
appears. Select
AS
in the
Controlled by
drop-down list.
Figure B-107. ICS-2000 Inject Valve Server Configuration Properties