ICS-2000 Ion Chromatography System
B-84
Doc. 031857-03 4/06
7.
Slide the EluGen cartridge down into the
holder until it stops (see
Make sure the cable remains out of the
holder.
NOTE To avoid crimping the lines, do not
twist the cartridge when you slide
it into the holder. The lines are
designed to coil inside the holder.
Figure B-81. EluGen Cartridge Installed in the Holder
8.
Remove the plug from
the
EGC-1
connector on
top of the ICS-2000.
9.
Orient the cartridge’s
blue 4-pin electrical
cable connector as
shown in
.
Push it onto the
EGC-1
connector on the ICS-
2000. Twist the ring on
the cable connector
fingertight to secure it.
Figure B-82. Connecting the EluGen Cartridge Cable
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...