2 • Features
Doc. 031857-03 4/06
2-31
under the timebase, and select
Edit>Properties
. Select the
Options
tab
Figure 2-16. Server Configuration Properties: Degas Mode Options
Degas Mode Options
•
Always Off
: The degas pump is always off.
•
Always On
: The degas pump is always on. This setting is for test
purposes by a Dionex Service Representative. Do not use this setting
for routine operation.
•
Cycle
: The degas pump cycles on and off according to the times
specified in the
Cycle On
and
Off
fields.
Cycle On
specifies for how
long the degas pump runs during a cycle.
Cycle Off
specifies the time
between cycles.
•
Monitor
: (default mode) The ICS-2000 monitors the degas pressure
reading and turns the degas pump on and off as required.