9. Disconnect the
EGC IN
and
EGC OUT
lines from the
INLET
and
OUTLET
ports on the cartridge.
10. Lift the EGC holder out
of the top cover and pull
the liquid lines through
the tubing chase (see
) to remove
them from the ICS-
2000.
Figure 5-70. Top Cover Tubing Chase
11. If a CR-TC is installed, go on to
.
12. If an EPM is installed, go on to
13. If neither a CR-TC or EPM is installed, go on to
.
Tubing Chase