5 • Service
Doc. 031857-03 4/06
5-57
Plumbing the EluGen Cartridge and EPM for Conditioning with
Carbonate/Bicarbonate Eluent
illustrates the liquid flow path through the eluent generator
during conditioning with carbonate/bicarbonate eluent.
Figure 5-52. Plumbing for Conditioning with Carbonate/Bicarbonate Eluent
EGC
IN
TO DEGAS
ELUENT OUT
TO SRS/AES
REGEN OUT
WASTE,
GAS
SEPARATOR
TO INJ
VALVE IN - P/
ELUENT OUT
EGC VENT
TO DEGAS
ELUENT IN
EGC
DEGAS
Connected to
Gas Separator
Assembly
Connected
Inject Valve
Port P (2)
to
Pre-plumbed
Pump Pulse
Damper
to
Inlet
Outlet
Send to waste
EGC CO
Mixer
3
EGC II
K CO
2
3
TO CR-TC/EPM
REGEN IN
(orange label)
TO PUMP/
DAMPER
TO CR-TC/EPM
ELUENT OUT
(yellow label)
TO
C
R-
TC
/E
PM
RE
GE
N
OU
T
(b
lue
la
be
l)
EPM
EL
UE
NT IN
T
O
C
R
-T
C
/E
P
M
E
L
U
E
N
T
IN
(r
ed
la
b
el
)
EGC
OUT
TO EGC CO3
MIXER IN
TO INJ VALVE
PORT P
TO EGC CO3
MIXER OUT
(yellow
label)
(blue
label)
(red
label)
Back-
pressure
Coil
10-32 to
1/4-28 Coupler
14 MPa (2000 psi)
at 1.0 mL/min
Temporarily
disconnect from
TO INJ VALVE IN -
P/ELUENT OUT
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...