ICS-2000 Ion Chromatography System
A-4
Doc. 031857-03 4/06
A.10 Column Heater
A.11 Eluent Generator
Operating
Temperature
30 to 60 °C (86 to 140 °F); must be at least 5 °C above the ambient
temperature
Concentration
Range
Dependent on the EluGen cartridge type, the cartridge configuration
(single or linked dual), and the flow rate. Refer to
and
.
EluGen Cartridge
Eluent Concentration Range
K
2
CO
3
0.1 to 15 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 2.0 mL/min flow
where X = 15/flow in mL/min
KOH
0.1 to 100 mM at
0.1
to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow in mL/min
LiOH
0.1 to 80 mM at 0.1 to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 80/flow in mL/min
MSA
0.1 to 100 mM at
0.1
to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow in mL/min
NaOH
0.1 to 100 mM at
0.1
to 1.0 mL/min flow
0.1 to X mM at 1.0 to 3.0 mL/min flow
where X = 100/flow mL/min
Table A-1. Eluent Concentration Ranges for Single-Cartridge Configurations