5 • Service
Doc. 031857-03 4/06
5-41
8. Disconnect the
EGC IN
and
EGC OUT
lines
from the
INLET
and
OUTLET
ports on the
EluGen cartridge (see
Figure 5-34. Disconnect the EluGen Cartridge Inlet and Outlet Port Lines
9. Follow the instructions in the EluGen cartridge manual, provided on
the Dionex Reference Library CD-ROM (P/N 053891), to prepare an
expended EluGen cartridge for disposal or to store a partially used
EluGen cartridge for later use.
10. Remove the new EluGen cartridge from the shipping box.
11. Orient the EluGen
cartridge with the fittings
facing up and remove the
plugs from the EluGen
cartridge
INLET
and
OUTLET
Figure 5-35. EluGen Cartridge Inlet and Outlet Ports
EGC OUT
OUTLET
Port
EGC IN
INLET
Port
OUTLET
INLET