Example AS40 Program Commands
AS40 Program Command Notes:
1. Close the ICS-2000
RELAY OUT 1
, which is connected to the AS40
LOAD
relay. This signals the AS40 to load the sample.
2. Zero the baseline.
3. Start data acquisition.
4. Switch the valve to Inject.
5. Stop data acquisition.
3.11 Process Samples
Samples can be run either manually (one at a time) or automatically (in groups or
“batches”).
3.11.1 Manual Sample Processing
To process samples manually, select operating parameters and commands
from the touch screen pages, or from the Chromeleon menu bar, toolbars,
and Control panel. Commands are executed as soon as they are entered.
NOTE When you run samples manually from the front panel
(without using
Chromeleon), the data is saved only
temporarily in the ICS-2000 memory. For a permanent
record of the data, connect a recording device (chart
recorder, integrator, etc.) to the analog output. See
for analog output connection instructions.
-2.300
Pump_ECD_Relay_1.Closed
Duration=138.00
;Note 1
0.000
Pump_ECD.Autozero
;Note 2
ECD_1.AcqOn
;Note 3
Pump_InjectValve.InjectPosition
Duration=30.00
;Note 4
30.00
ECD_1.AcqOff
;Note 5
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...