C • Touch Screen Operation
Doc. 031857-03 4/06
C-7
2.00 mL/min. Refer to the column manual for the flow rate to use with your
column and application. Column manuals are included on the Dionex Reference
Library CD-ROM (P/N 053891).
Pump On/Off Control
Touching the
On
button immediately starts the pump flow at the rate displayed in
the
Flow Rate
field. Touching
Off
stops the flow.
Prime
The
Prime
button is reserved for use during the pump head priming procedure.
Touching the
Prime
button displays a confirmation page that reminds you to open
the waste valve. Open the valve, and then touch the
OK
button to begin priming.
Or, to exit the page without priming, touch the
Cancel
button. To stop priming,
touch the
Off
button. During priming, the flow rate is approximately 3 mL/min.
See
for detailed priming instructions.
Eluent Level
Use this field to monitor the amount of liquid in the eluent reservoir. When you
fill the reservoir, enter the amount of liquid in the reservoir. For example, if the
reservoir contains 200 mL and you add 1 liter, enter 1.2 liters. Thereafter, the ICS-
2000 determines the eluent usage by monitoring the flow rate and the length of
time the pump is on. The ICS-2000 updates the
Eluent Level
field as the eluent is
used up. A warning first appears if the level falls below 200 mL. Warnings are
repeated at 100 mL and 0 mL.
In order for the eluent level displayed on the screen to be accurate,
the level must be entered by the user when the reservoir is filled. The
ICS-2000 does not automatically detect when the reservoir is filled,
nor when it is empty.
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...