ICS-2000 Ion Chromatography System
5-28
Doc. 031857-03 4/06
5.10 Replacing the Suppressor
Refer to the suppressor manual for guidance about when to replace a suppressor.
Suppressor manuals are included on the Dionex Reference Library CD-ROM
(P/N 053891).
1. Refer to the suppressor quick start guide (shipped with the suppressor) for
instructions on preparing the suppressor before initial use and for additional
installation details.
2. Turn off the pump from the ICS-2000 Control panel in Chromeleon or from
the touch screen
HOME
page.
3. Open the front door of the ICS-2000.
4. Disconnect the two eluent and the two regenerant lines from the suppressor.
5. To remove the suppressor from the component mounting panel, slide it up to
detach it from the mounting tabs on the panel and then pull it toward you.
6. Continue pulling the
suppressor out until the
yellow connector on the
cable inside the ICS-2000 is
outside the panel (see
7. Disconnect the suppressor
cable (twist the ring on the
yellow connector and pull
the two connectors apart).
Figure 5-19. Removing the Suppressor
8. Connect the cable from the new suppressor to the ICS-2000 cable (align the
pins on the two connectors and push them together).
9. Orient the suppressor with the
REGEN OUT
fitting on top and the cables to the
right.
Yellow
Connector
Содержание Dionex ICS-2000
Страница 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Страница 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Страница 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Страница 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Страница 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Страница 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...