5 • Service
Doc. 031857-03 4/06
5-33
tubing) and press the heat exchanger onto the heater. Replace the screws
removed in
.
10. Connect the new heat exchanger inlet line to Port C on the injection valve and
connect the outlet line to the guard column inlet.
11. Reinstall the top metal plate and the heater cover.
5.13 Replacing the Eluent Valve
1. Turn off the power to the ICS-2000.
2. Open the ICS-2000 front door.
3. To prevent an eluent leak during the valve replacement procedure, disconnect
the eluent line from the reservoir cap. Or, plug the line on the left side of the
valve with a coupler (P/N 039056 and a plug (P/N 037268) after
disconnecting the liquid line from the valve
.
4. Disconnect the two liquid
lines connected to the
eluent valve (see
5. Unscrew and remove the
two screws that attach the
valve to the component
panel (see
Save the screws.
Figure 5-24. Eluent Valve
Disconnect
Remove